Serial Dilution Calculator
Each tube is made by adding the transfer volume from the previous tube (tube 1 from the stock) to the diluent, then mixing well. After you take the next transfer out, each tube keeps its volume minus that transfer.
How a serial dilution works
A serial dilution repeats the same dilution several times, each tube made from the one before. With a dilution factor DF per step, tube n has concentration:
Cn = C0 ÷ DFn
To make each tube with a total volume V: transfer = V ÷ DF and diluent = V − transfer. Splitting a big dilution into steps is far more accurate than one giant step: pipetting 1 µL into 999 µL carries much more error than three 1:10 steps.
Worked example
A 1 mg/mL stock, five 1:10 steps, 1,000 µL per tube: transfer 100 µL into 900 µL of diluent each time. The tubes hold 0.1, 0.01, 0.001, 0.0001 and 0.00001 mg/mL, a total dilution of 1:100,000 in tube 5.
A 1:2 series (DF = 2) halves the concentration each step: 500 µL into 500 µL. It's common for standard curves and antibiotic susceptibility tests.
Common mistakes
- Not mixing between steps. Mix each tube thoroughly before the next transfer, or every later tube is off.
- Reusing the same tip. Carry-over from a concentrated tube contaminates the next. Change tips each step.
- Confusing dilution factor with ratio. 1:10 means 1 part sample in 10 parts total (1 + 9), not 1 + 10.
- Forgetting the volume you remove. Each tube loses the transfer volume when you make the next one; plan your volumes so the last tubes still have enough.
FAQ
How is this different from a single dilution?
A single dilution uses C1V1 = C2V2 once; see the dilution calculator. A serial dilution chains several.
Can I pipette mL volumes quickly?
For 1–80 mL transfers with serological pipettes, an electric pipette filler speeds things up and avoids mouth pipetting.
How do I make the stock in the first place?
Weigh the solute with the molarity calculator and an analytical balance for small masses.
More tools: free lab calculators.